Remote Control
R&S
®
FSPN
357
User Manual 1179.4363.02 ─ 01
[SENSe:]ADJust:CONFigure:FREQuency:LIMit:HIGH
<Frequency>
This command defines the upper limit of the frequency search range.
Prerequisites for this command
●
Turn on the automatic signal search with
Parameters:
<Frequency>
<numeric value>
Range:
See data sheet
*RST:
8 GHz
Default unit: Hz
Example:
//Define a frequency search range from 10 MHz to 500 MHz
ADJ:CONF:FREQ:LIM:LOW 10MHZ
ADJ:CONF:FREQ:LIM:HIGH 500MHZ
Manual operation:
See
[SENSe:]ADJust:CONFigure:FREQuency:LIMit:LOW
<Frequency>
This command defines the lower limit of the frequency search range.
Prerequisites for this command
●
Turn on the automatic signal search with
Parameters:
<Frequency>
<numeric value>
Range:
See data sheet
*RST:
1 MHz
Default unit: Hz
Example:
//Define a frequency search range from 10 MHz to 500 MHz
ADJ:CONF:FREQ:LIM:LOW 10MHZ
ADJ:CONF:FREQ:LIM:UPP 500MHZ
Manual operation:
See
[SENSe:]FILTer:LPASs:FREQuency:MANual
<Frequency>
This command defines the cutoff frequency of the low pass filter you can use to mea-
sure small carrier frequencies.
Prerequisites for this command
●
Turn on low pass filter (
Parameters:
<Frequency>
<numeric value>
*RST:
1 MHz
Default unit: Hz
Common Measurement Settings