Remote Control
R&S
®
FSPN
481
User Manual 1179.4363.02 ─ 01
Parameters:
<State>
ON | OFF | 1 | 0
*RST:
0
Example:
ROSC:OSYN ON
Manual operation:
See
[SENSe:]ROSCillator:SOURce
<Source>
This command selects the reference oscillator.
If you want to select the external reference, it must be connected to the R&S
FSPN.
Parameters:
<Source>
INTernal
The internal reference is used (10 MHz)
EXTernal | EXTernal1 | EXT1
The external reference from the "REF INPUT 10
MHZ" connec-
tor is used; if none is available, an error flag is displayed in the
status bar
EAUTo
The external reference is used as long as it is available, then the
instrument switches to the internal reference
Example:
ROSC:SOUR EXT
Manual operation:
See
[SENSe:]ROSCillator:SOURce:EAUTo?
This command queries the current reference type in case you have activated an auto-
matic switch to the internal reference if the external reference is missing.
Return values:
<Reference>
INT | EXT
INT
internal reference
EXT
external reference
Example:
SENS:ROSC:SOUR:EAUT?
Queries the currently available reference type.
Usage:
Query only
Configuring the R&S
FSPN