Remote Control
R&S
®
FSPN
334
User Manual 1179.4363.02 ─ 01
Suffix:
<n>
.
1..n
Query parameters:
<Trace>
TRACE1 | ... | TRACE6
Queries the data of one of the colored line traces.
XGINdicator
Queries the data of the cross-correlation gain indicator (grey
area).
<Points>
<numeric value> (integer only)
Defines the number of linearily spaced measurement points that
are returned within the specified frequency range (defined by the
start and stop values).
<Start>
<numeric value>
Defines the start offset of the frequency range whose trace data
you like to query.
The start offset must be within the measurement range you have
defined. If you do not define a start and stop offset, the com-
mand returns the trace data for the complete measurement
range that was defined.
Minimum start offset is 0 Hz.
Default unit: Hz
<Stop>
<numeric value>
Defines the stop offset of the frequency range whose trace data
you like to query.
The stop offset must be within the measurement range you have
defined. If you do not define a start and stop offset, the com-
mand returns the trace data for the complete measurement
range that was defined.
Maximum stop offset depends on the measurement range.
Default unit: Hz
Return values:
<Result>
String that contains the trace data as comma-separated values.
Each data point consists of two values: offset frequency in Hz
and level in dBc/Hz.
The number of value pairs depends on the number of points you
have defined.
Example:
//Query linear trace data
TRAC:LIN? TRACE1,5,100,100000
would return, e.g.
100.000000,0.000000,25075.000000,-139.424408,
50050.000000,-140.380112,75025.000000,
-139.754211,100000.000000,-141.686478
Usage:
Query only
Measurements and Result Displays