MAINTENANCE
6-7
6.3
Sampling Pump
When approaching the end of the specified lifetime of the
pump, it will consume higher amount of energy and reduce its
sample draw capability significantly. When this occurs, it is
necessary to replace or rebuild the pump. When checking the
pump flow, make sure that the inlet connector is tight and the
inlet tubing is in good condition. Connect a flow meter to the
gas inlet probe. The flow rate should be above 450 cc/min when
there is no air leakage.