PROGRAMMING
4-8
4.4.1
Fresh Air Calibration
This procedure determines the zero point of the sensor
calibration curve. To perform a fresh air calibration, use the
calibration adapter to connect the MiniRAE 2000 to a “fresh”
air source such as from a cylinder or Tedlar bag (option
accessory). The “fresh” air is clean dry air without any organic
impurities. If such an air cylinder is not available, any clean
ambient air without detectable contaminant or a charcoal filter
can be used.
1. The first sub-menu shows: “Fresh air Cal?”
2.
Make sure that the MiniRAE 2000 is connected to one of
the “fresh” air sources described above.
3.
Press the
[Y/+]
key, the display shows “zero in progress”
followed by “wait..” and a countdown timer.
After about 15 seconds pause, the display will show the
message “update data…zeroed… reading = X.X ppm…” Press
any key or wait about 20 seconds, the monitor will return back
to “Fresh air Calibration?” submenu.