![Rae MiniRAE 2000 PGM-7600 Operation And Maintenance Manual Download Page 78](http://html1.mh-extra.com/html/rae/minirae-2000-pgm-7600/minirae-2000-pgm-7600_operation-and-maintenance-manual_781222078.webp)
THEORY OF OPERATION
5-1
5.
THEORY OF OPERATION
The MiniRAE 2000 monitor uses a newly developed
electrodeless discharge UV lamp as the high-energy photon
source for the PID. As organic vapors pass by the lamp, they are
photo-ionized and the ejected electrons are detected as a
current. The PID sensor with a standard 10.6 eV lamp detects a
broad range of organic vapors. A lamp with high photon energy
(e.g. 11.7 eV) will measure the more kinds of compounds,
whereas low photon energies (e.g. 9.8 eV) are selective for
easily ionizable compounds such as aromatics. In principle, any
compound with an ionization energy lower than that of the lamp
photons can be measured.
The PID sensor for the MiniRAE 2000 monitor is constructed
as a small cavity in front of the UV lamp. A diaphragm pump
draws the gas sample into the sensor and then pumps it out
through the side of the instrument.
Figure 5-1 MiniRAE 2000 Block Diagram
Micro-
processor
LCD Display
Keypad
Serial
Port
Battery
PID Sensor
UV lamp
Pump
Light
sensor
LED
Buzzer
Charger
Sample Gas
Inlet Probe
Sample Gas
Outlet
Summary of Contents for MiniRAE 2000 PGM-7600
Page 93: ......