PROGRAMMING
4-13
4.4.5
Modify Cal Memory
If the current cal memory number selected is not memory 0,
users will be prompted whether to modify the settings of the
selected cal memory. Press
[Y/+]
to modify the cal memory and
[N/-]
to go to the next sub-menu.
Once
[Y/+]
is pressed the LCD display will show the current
memory number, current Gas selected and prompt user for
acceptance of current gas selected.
1.
Press
[N/-]
to modify the gas selection if desired. Or press
[Y/+]
key to skip the change of gas selection, and proceed to
the next sub-menu.
2.
After pressing
[N/-]
, display shows “Copy gas from
library?” Press
[Y/+]
to accept or
[N/-]
for the next sub-
menu, “Enter Custom gas?”
3.
In the “Copy gas from library” submenu, use
[Y/+]
and
[N/-
]
keys to scroll through the selections in the library. Press
[MODE]
key momentarily to select the gas. The display
shows ”Save?” Press
[Y/+]
to save or
[N/-]
to discard the
changes and proceed to next sub-menu.
4.
In the Custom gas sub-menu, the user can enter the gas
name. Press the
[Y/+]
or
[N/-]
key to cycle through all 26
letters and 10 numerals. Press the
[MODE]
key
momentarily to advance to the next digit. The flashing digit
will move to the next digit to the right. Repeat this process
until all digits (up to 8 digits) of the custom gas name is
entered.
Press and hold the
[MODE]
key for 1 second to exit the name
entry mode. The display will show “Save?“ Press
[Y/+]
to save
the entry, or
[N/-]
to discard the changes.
Summary of Contents for MiniRAE 2000 PGM-7600