![Rae MiniRAE 2000 PGM-7600 Operation And Maintenance Manual Download Page 21](http://html1.mh-extra.com/html/rae/minirae-2000-pgm-7600/minirae-2000-pgm-7600_operation-and-maintenance-manual_781222021.webp)
OPERATION
2-10
Note: This display appears only after a measurement has
been started. It does not appear when the user is cyling
through the Main Menu and the monitor is idling.
•
x.x ppm:
(read “Taking a Measurement” for details)
Note: This display appears only after a measurement has
been started. It does not appear when the user is cyling
through the Main Menu and the monitor is idling.
•
TWA/Avg:
Displays (in ppm) the Time Weighted Average
(TWA) or the Average since the start of the measurement.
The average is recalculated every minute.
•
STEL:
Displays the Short Term Exposure Limit.
•
PEAK:
Displays (in ppm) the highest instantaneous reading
since the start of the measurement. If
[Y/+]
is pressed while
the peak reading is displayed, the unit will ask
Reset Peak?
.
If
[Y/+]
is pressed again, the peak value will be cleared and
the display will return to the
Ready…
message or
instantaneous reading. The peak reading is automatically
reset when a new measurement is started by pressing
[Y/+]
from the
Ready…
screen.
•
Run time hh:mm:
The duration of the current
measurement period.
•
Bat = X.XV:
The current battery voltage.
Note: A fully charged battery pack should show 4.8 volts or
higher. When the battery voltage falls below 4.4 volts, a
flashing
“Bat”
will appear as a warning message. At that
point, you have 20-30 minutes of run time remaining. When
the battery voltage falls below 4.2 volts the monitor turns
off automatically.
•
Mm/dd/yy:
The current date.
•
hh:mm:
The current time (24-hour format)