OPERATION OF ACCESSORIES
3-5
3.2
Optional Accessories
1) Dilution Fitting
The user may wish to install a dilution fitting on the inlet to
dilute the gas samples. One application for a dilution fitting is to
measure organic gas when the concentration exceeds the upper
limit of the sensor range.
Make sure to set the dilution ratio in the programming mode
(see Section 4.7.9) so that the correct gas reading will be
displayed when the dilution fitting is used.
WARNING: To use a dilution fitting, the user must have
the monitor located in a clean atmosphere outside the
confined space and use a remote access probe or Tygon
tubing to measure the gas concentration inside the confined
space.
2) Calibration Adapter
The calibration adapter for the MiniRAE 2000 is a simple 6-
inch Tygon tubing with a metal adapter on one end. During
calibration, simply insert the metal adapter into the regular gas
inlet probe of the MiniRAE 2000 and the tubing to the gas
regulator on the gas bottle.