PROGRAMMING
4-12
4.4.4
Change Span Value
This function allows the user to change the span values of the
calibration gases.
1.
“Change Span Value?” is the fourth sub-menu item in the
Calibration sub-menu
2.
Press
[Y/+]
, display shows the gas name and the span value.
A cursor will blink at the first digit of the Span value. To
modify the span gas value, go to Step 3. Otherwise, press
and hold the
[MODE]
key for 1 second to accept the
previously stored span gas value and move to the next sub-
menu.
3.
Starting from the left-most digit of the span gas value, use
the
[Y/+]
or
[N/-]
key to change the digit value and press
[MODE]
key momentarily to advance to next digit. Repeat
this process until all digits are entered. Press and hold the
[MODE]
for 1 second to exit.
4.
The display shows “Save?” To accept the new value, press
the
[Y/+]
key. Press the
[N/-]
key or the
[MODE]
key to
discard the change and move to the next sub-menu.