![Rae MiniRAE 2000 PGM-7600 Operation And Maintenance Manual Download Page 64](http://html1.mh-extra.com/html/rae/minirae-2000-pgm-7600/minirae-2000-pgm-7600_operation-and-maintenance-manual_781222064.webp)
PROGRAMMING
4-26
4.7.1
Change Operation Mode
MiniRAE 2000 supports two operation modes: Survey and
Hygiene mode.
Survey mode:
Manual start/stop of measurements and display
of certain exposure values.
Hygiene mode:
Automatic measurements, running and
datalogging continuously and calculates additional exposure
values.
1. “Change Op Mode?” is the first sub-menu item in the
Monitor Setup menu (Table 4.7).
2.
Press the
[Y/+]
key and the display shows the current user
mode: “Op Mode =
current mode
?”
3.
Press the
[Y/+]
key to accept the currently displayed
operation (Op) mode. Press
[N/-]
to toggle to the other
operation mode. Press
[MODE]
to exit this sub-menu and
move to the next monitor setup sub-menu.
4.
When changing Op mode from Hygiene to Survey, the
display shows the additional message “Warning! Exit
Hygiene?” to prevent accidental exit from Hygiene mode.
Press the
[Y/+]
key to acknowledge.
5.
If there is any change to the existing setting, the display will
show “Save?” Press the
[Y/+]
key to accept or the
[N/-]
key
to discard and move to the next sub-menu.
Note
: If a new Op Mode is saved, the display shows “Op Mode
changed!!” when exiting the programming mode.
Summary of Contents for MiniRAE 2000 PGM-7600