OPERATION
2-19
2.7
Integrated Sampling Pump
The MiniRAE 2000 portable VOC monitor includes an
integrated sampling pump. This is a diaphragm type pump that
provides a 500-600 cc per minute flow rate. Connecting a
Teflon or metal tubing with 1/8 inch inside diameter to the gas
inlet port of the MiniRAE 2000, this pump can pull in air
samples from 200 feet away horizontally, or 90 feet vertically,
at about 3 feet per second flow speed.
The pump turns on when a measurement is started, and turns off
when the sample is manually stopped in Survey mode or when
the unit is turned off from Hygiene Mode.
If liquid or other objects are pulled into the inlet port filter, the
monitor will detect the obstruction and shut down the pump
immediately. The alarm will be activated and a flashing error
message “Pump” will be also displayed on the LCD display.
The user should acknowledge the pump shut off condition by
clearing the obstruction and pressing the
[Y/+]
key to re-start
the pump.
The pump stall threshold is set in the special Diagnostic Mode
(Section 8).