MAINTENANCE
6-6
To clean lamp housing or change the lamp:
To clean lamp housing or change the lamp:
If the lamp does not turn on, the monitor will display an error
message to indicate replacement of the lamp may be required.
1. If the lamp is operational, clean the lamp window surface
and the lamp housing by wiping it with GC grade methanol
using a cotton swab using moderate pressure. After
cleaning, hold the lamp up to the light at an angle to detect
any remaining film. Repeat the process until the lamp
window is clean. Never use water solutions to clean the
lamp. Dry the lamp and the lamp housing thoroughly after
cleaning.
CAUTION: Never touch the window surface with
the fingers or anything else that may leave a film.
Never use acetone or aqueous solutions.
2. If the lamp does not turn on, remove the lamp from the lamp
housing. Place the lamp O-ring onto the new lamp. Insert
the new lamp, avoiding contact with the flat window
surface.
3. Reinstall the PID sensor module.
4. Tighten the Lamp Housing Cap.
5. If the lamp type has been changed, adjust the lamp type
setting in the programming mode (Section 4.7.8).