OPERATION OF ACCESSORIES
3-6
3) Calibration Regulator and Flow Controller
The Calibration Regulator and Flow controller is used in the
calibration process. It regulates the gas flow rate from the Span
gas cylinder into the gas inlet of the MiniRAE 2000 monitor
during calibration process. The maximum flow rate allowed by
the flow controller is about 0.5L/min (500 cc per min.).
Alternatively, a Demand-flow Regulator or a Tedlar gas bag
may be used to match the pump flow precisely.
4)
Organic Vapor Zeroing kit (Charcoal filter)
The Organic Vapor Zeroing Kit is used for filtering organic air
contaminants that may affect the zero calibration reading. To
use the Organic Vapor Zeroing Kit, simply connect the filter to
the inlet port of the MiniRAE 2000.