PROGRAMMING
4-11
4.4.3
Select Cal Memory
This function allows the user to select one of eight different
memories for gas calibration and measurement. For Memories
#1-7, the calibration and measurement gas is the same and no
correction factor is applied. For Memory #0, the calibration gas
is always isobutylene and the measurement gas may be
different, in which case the correction factor for that gas is
automatically applied. The default gas selections are listed in
Section 4.4
1.
“Select Cal Memory?” is the third sub-menu item in the
Calibration sub-menu. Pressing the
[Y/+]
key, the display
will show “Gas =” gas name followed by “Mem # x?”
2.
Press
[N/-]
to scroll through all the memory numbers and
the gas selections respectively. Press
[Y/+]
to accept the
displayed Cal Memory number.
3.
After the
[Y/+]
key is pressed, the display shows “Save?”
Press
[Y/+]
key to save and proceed. Press
[N/-]
to discard
the entry and advance to the next sub-menu.
4.
If the gas in a newly selected Cal Memory number is not
calibrated, the display shows “CF= x.xx”. A correction
factor with the value “x.xx” will be applied.
5.
If the gas of a newly selected cal memory number has been
calibrated previously, the display shows “Last calibrated
xx/xx/xx”.