GENERAL INFORMATION
1-3
Direct Readout:
Instantaneous, average, STEL and peak value,
battery voltage and elapsed time
Intrinsic Safety:
UL & cUL Class 1, Division I, Group A,B,C,D,
Temperature Code T3C (US & Canada);
0575
II 1G DEMKO 02 ATEX 0204759
Eex ia IIC T4 (Europe)
EM Interference:
No effect when exposed to 0.43 W/cm2 RF
interference (5 watt transmitter at 12 inches)
Alarm Setting:
Separate alarm limit settings for Low, High,
STEL and TWA alarm
Operating Mode:
Survey or Hygiene mode
Alarm: 90 dB buzzer and flashing red LEDs to indicate exceeded preset
limits, low battery voltage, or sensor failure.
External Alarm:
Optional plug-in pen-size vibration alarm or
remote alarm
Alarm Mode:
Latching or automatic reset
Real-time Clock:
Automatic date and time stamps on data logged
information
Datalogging:
15,000 points with time stamp, serial number,
user ID, site ID, etc.
Communication:
Upload data to PC and download instrument setup
from PC through RS-232 port
Sampling Pump:
Internally integrated. Flow rate: 450-550 cc/min.
Temperature:
0º to 45ºC (32º to 113ºF)
Humidity:
0 % to 95 % relative humidity
(non-condensing)
Housing:
ABS + PC, conductive coating, splash and dust
proof, will withstand 1 meter drop test with
rubber boot
Attachment:
Wrist strap, rubber boot and belt clip