![Racal Instruments 3162A/B User Manual Download Page 92](http://html1.mh-extra.com/html/racal-instruments/3162a-b/3162a-b_user-manual_3132596092.webp)
User Manual 3162A/B
SCPI Command Reference 4-14
Modulation
Command
Summary
The Modulation Commands controls amplitude and phase modulation
parameters. Optional modes may be omitted from these commands. Factory
defaults after *RST are shown in bold typeface. Parameter low and high limits
are given where applicable.
Command and Parameters
[SOURce:]
AM {
OFF
| ON}
AM?
PM {
OFF
| ON}
PM?
Trigger Command
Summary
The Trigger commands control the trigger modes of the Model 3162B. The
Model 3162B can be placed in Triggered, Gated or Burst mode. Trigger source
is selectable from an external source, internal trigger generator, backplane
TTLTrg 0-3, ECLTrg0, or a software trigger. Optional modes are omitted from
these commands. Factory defaults after *RST are shown in bold typeface.
Parameter low and high limits are given where applicable
.
Command and Parameters
(Default,
Low
Limit,High
Limit
)
:INITiate:CONTinuous {OFF |
ON
}
:TRIGger:BURSt {
OFF
| ON}
:TRIGger:COUNt <value>
(
1
,1,1e6)
:TRIGger:DELay <value>
(
0
,0,8e6)
:TRIGger:DELay:STATe {
OFF
| ON}
:TRIGger:GATE {
OFF
| ON}
:TRIGger:LEVel <value>
(
1.6
,-10,10)
:TRIGger:SLOPe {
POSitive
| NEGative}
:TRIGger:SOURce:ADVance {
EXTernal
| INTernal | TTLTrg<n> | ECLTrg0}
:TRIGger:TIMer: <value>
(
100e-6
,20e-6,1000)
:TRIGger:IMMediate
*TRG
PLL Command
Summary
PLL commands control the PLL mode and the phase offsets versus another
reference. The reference signal is applied either to a front panel connector or
through one of the VXIbus trigger lines. The Model 3162B in PLL mode, can be
used as a slave only. After lock, the generator’s output frequency is exactly
equal to the frequency of the reference signal. Note that the PLL lock range
depends on the number of waveform samples, sampling frequency and external
reference frequency. The relationship between these parameters is shown in
Appendix A. Also included in this section are phase offset command. The
Revision 4/18/00
Summary of Contents for 3162A/B
Page 4: ...This page was left intentionally blank...
Page 10: ...3162A B User Manual vi Support Offices 6 2...
Page 12: ...3162A B User Manual viii This page was left intentionally blank...
Page 17: ...User Manual 3162A B Getting Started 1 5 Figure 1 2 The Front Panel...
Page 23: ...User Manual 3162A B Getting Started 1 11 Figure 1 7 WaveCAD Example...
Page 27: ...User Manual 3162A B Getting Started 1 15 This page was left intentionally blank...
Page 28: ......
Page 34: ...3162A B User Manual Configuring The Instrument 2 6 This page was left intentionally blank...
Page 78: ......