User Manual 3162A/B
Maintenance and Performance Checks 5-5
Frequency Accuracy
Accuracy Specifications:
1 ppm of reading to 200.00 MHz
Equipment:
Counter
1.
Configure the Model 3162B as follows:
Function:
Square
Frequency: 10 Hz
Amplitude:
1 V
Remote Commands: :RES
:OUTP ON
:FUNC:SHAP SQU
:FREQ 10
:VOLT 1
2.
Set the counter to frequency measurement.
3.
Connect the Model 3162B output to the counter’s input. Change the
frequency setting as required for the test and verify the reading on
the counter display as follows:
3162B Setting
Counter Reading
10.00000 Hz
9.999999 Hz - 10.00001 Hz
100.0000 Hz
99.99999 Hz - 100.0001 Hz
1.000000 kHz
999.9999 Hz - 1.000001 kHz
10.00000 kHz
9.999999 kHz - 10.00001 kHz
100.0000 kHz
99.99999 kHz - 100.0001 kHz
1.000000 MHz 999.9999 kHz - 1.000001 MHz
10.00000 MHz 9.999999 MHz - 10.00001 MHz
150.0000 MHz 149.9998 MHz - 150.0002 MHz
200.0000 MHz 199.9998 MHz - 200.0002 MHz