![Racal Instruments 3162A/B User Manual Download Page 111](http://html1.mh-extra.com/html/racal-instruments/3162a-b/3162a-b_user-manual_3132596111.webp)
User Manual 3162A/B
Maintenance and Performance Checks 5-9
Square Wave
Characteristics
Specified Transition Time
: <2.5 ns typical (10% to 90% of amplitude)
Specified Aberration
: <5% of amp10 mV
Equipment:
Oscilloscope, 20dB attenuator
1.
Configure the Model 3162B as follows:
Function:
Square wave
Frequency:
10 MHz
Amplitude:
3 V
Remote Commands:
:RES
:FUNC:SHAP SQU
:FREQ 10e6
:VOLT 3
:OUTP ON
2.
Connect the A channel output to the oscilloscope input. Use the
20dB attenuator and set oscilloscope input impedance to 50
Ω
.
3
.
Set the oscilloscope and verify that the rise and fall times are less
than 2.5 ns. Verify that overshoot and ringing are less than 5% of
amplitude.
4.
Configure Channel B using the following commands:
Remote Commands:
:INST:SEL B
:FUNC:SHAP SQU
:VOLT 3
:OUTP ON
5.
Connect the B channel output to the oscilloscope input. Use the
20dB attenuator and set oscilloscope input impedance to 50
Ω
.
6.
Set the oscilloscope and verify that the rise and fall times are less
than 2.5 ns. Verify that overshoot and ringing are less than 5% of
amplitude.
Summary of Contents for 3162A/B
Page 4: ...This page was left intentionally blank...
Page 10: ...3162A B User Manual vi Support Offices 6 2...
Page 12: ...3162A B User Manual viii This page was left intentionally blank...
Page 17: ...User Manual 3162A B Getting Started 1 5 Figure 1 2 The Front Panel...
Page 23: ...User Manual 3162A B Getting Started 1 11 Figure 1 7 WaveCAD Example...
Page 27: ...User Manual 3162A B Getting Started 1 15 This page was left intentionally blank...
Page 28: ......
Page 34: ...3162A B User Manual Configuring The Instrument 2 6 This page was left intentionally blank...
Page 78: ......