User Manual 3162A/B
Maintenance and Performance Checks 5-21
Offset Adjustment
Equipment:
DMM, 50
Ω
feedthrough termination
1.
Configure the Model 3162B as follows:
Function:
Sine
Frequency: 1 MHz
Amplitude:
10 mV
Offset:
+4.500 V
Remote Commands: :RES
:FREQ 1e6
:VOLT 10e-3
:VOLT:OFFS 4.5
:OUTP ON
Channel A Adjustment
2.
Connect the Model 3162B output to the DMM input through the 50
Ω
feedthrough termination. Set the DMM to DCV measurements and 10 V
range. Record this reading.
3.
Change the offset setting to -4.50 V and adjust R95 on the main board
to have the exact reading as was recorded in the previous step (with
reversed polarity).
4.
Adjust R96 on the main board for a DMM reading of -4.500 V, ±30
mV.
5.
Repeat Steps 3 and 4 until the best result is obtained in both steps.
Channel B Adjustment
6.
Change the Model 3162B settings using the following remote
commands:
:INST:SEL B
:VOLT 10e-3
:VOLT:OFFS 4.5
:OUTP ON
7.
Connect the Model 3162B output to the DMM input through the 50
Ω
feedthrough termination. Set the DMM to DCV measurements and 10 V
range. Record this reading.
8.
Change the offset setting to -4.50 V and adjust R77 on the B channel
engine board to have the exact reading as was recorded in the previous
step (with reversed polarity).
9.
Adjust engine board B R76 until the DMM reading is -4.500 V, ±30 mV.
10. Repeat steps 3 and 4 until the best result is obtained in both steps.
Summary of Contents for 3162A/B
Page 4: ...This page was left intentionally blank...
Page 10: ...3162A B User Manual vi Support Offices 6 2...
Page 12: ...3162A B User Manual viii This page was left intentionally blank...
Page 17: ...User Manual 3162A B Getting Started 1 5 Figure 1 2 The Front Panel...
Page 23: ...User Manual 3162A B Getting Started 1 11 Figure 1 7 WaveCAD Example...
Page 27: ...User Manual 3162A B Getting Started 1 15 This page was left intentionally blank...
Page 28: ......
Page 34: ...3162A B User Manual Configuring The Instrument 2 6 This page was left intentionally blank...
Page 78: ......