User Manual 3162A/B
Maintenance and Performance Checks 5-4
Initial Instrument
Setting
To avoid confusion as to which initial setting is to be used for each test, it is
required that the instrument be reset to factory default values prior to each test.
To reset the Model 3162A/B to factory defaults, send either of the following
commands:
RESet, or *RST.
Recommended Test
Equipment
Recommended test equipment for troubleshooting, calibration and performance
checking is listed below. Test instruments other than those listed may be used
only if their specifications equal or exceed the required characteristics.
Test Instrument & Accessories
Minimum Specification
Universal Counter
300MHz, 1 ppm accuracy
DMM
ACV, 0.2%; DCV, 0.1%
accuracy
Distortion Analyzer
10Hz to 100kHz
Spectrum Analyzer
1 GHz bandwidth, 1kHz
resolution
Pulse/Function Generator
50MHz, 10ns pulse width
Oscilloscope
2 channels, 500MHz analog
bandwidth
Feedthrough termination
50
Ω
, 0.1%
BNC Cables
1 meter long, 50
Ω
impedance
20dB feedthrough attenuator
50
Ω
, 1%
“T” BNC Connector
50
Ω
, 1%
Performance Check
Procedures
Use the following procedures to check the Model 3162A and the 3162B
against the specifications. A complete set of specifications is listed in
Appendix A. The following paragraphs show how to set up the instrument
for the test, what the specifications for the tested function are, and what
acceptable limits for the test are. If the instrument fails to perform within
the specified limits, the instrument must be calibrated or tested to find the
source of the problem.
In the following, 3162B tests refer to both 3162A and 3162B
models. If you own the Model 3162A, simply ignore tests that
pertain to the B channel operation.
Summary of Contents for 3162A/B
Page 4: ...This page was left intentionally blank...
Page 10: ...3162A B User Manual vi Support Offices 6 2...
Page 12: ...3162A B User Manual viii This page was left intentionally blank...
Page 17: ...User Manual 3162A B Getting Started 1 5 Figure 1 2 The Front Panel...
Page 23: ...User Manual 3162A B Getting Started 1 11 Figure 1 7 WaveCAD Example...
Page 27: ...User Manual 3162A B Getting Started 1 15 This page was left intentionally blank...
Page 28: ......
Page 34: ...3162A B User Manual Configuring The Instrument 2 6 This page was left intentionally blank...
Page 78: ......