Troubleshooting
Ozone Generator
Air Preparation – Heat Regenerative Dry Air
Problem/Symptom
Possible Cause
Solution
Air Prep LED not flashing
-Air dryer board not functioning
-Replace air dryer board
Dryer 1 or 2 LED not
illuminated
-Air dryer board not functioning
-Dryer 1 LED will not illuminate when
Dryer 2 LED is illuminated
-Dryer 1 is in cool down mode
-Dryer 2 LED will not illuminate when
Dryer 1 LED is illuminated
-Dryer 2 LED is in cool down mode
-Replace air dryer board
-See “Theory of Operation and Product
Description - Air Preparation System”
-See “Ozone Generator LED Display, Figure
11”
Dryer chamber(s) not
heating
-Heating element not functioning
-Replace Heating element
Indicating desiccant
cartridge has changed from
blue & white to all pink or
white. Moisture has entered
air prep system.
-Unit does not have constant power
-Excessive duty cycle
-Excessive relative humidity
-Solenoid valve not operating
-Air dryer board not functioning
-Unit must have constant power
-Duty cycle must not exceed 10 hours in a 24
hour period
-Relative humidity must not exceed 75%
-Replace solenoid valve
-Replace air dryer board
Ozone Generator
Problem/Symptom Possible Cause
Solution
Low vacuum
-Hydraulics/Pneumatics out of adjustment
-Defective check valve(s)
-Loose internal fittings
-Defective dielectrics
-See “Reclaim Manual – Low Vacuum Fault”
-Replace check valves
-Check all fittings, tighten as required
-Check & replace as required
Unit flooded with water
-Defective check valve(s)
-Assess damage, repair as required, replace
check valve(s)
Ozone smell detected from
or near ozone generator
-Insufficient vacuum at venturi
-Loose internal fittings
-Defective O-ring seals in reaction
chamber(s)
- See “Reclaim Manual – Low Vacuum Fault”
-Check all fittings, tighten as required
-Check & replace as required