System Troubleshooting - Minor Fault
Ozone Fault
“Touch” the Yellow Fault Box
to go to the next screen. Then
“touch” the “Internal Fault”
button on that screen to arrive at
the bottom screen.
If the system is displaying an
“Internal Fault”, it can be one of
several different components
inside of the ozone generator
itself. It’s a good idea to do
visual checks of the 48Vdc
Power Supply, the drive
board(s), the transformer(s),
and the reaction chamber(s)
also known as cells. See the
following pages. Internal faults
can be a bit challenging to
isolate.
Check the frint of the Ozone
Generator Cabinet & look at the
digital message display. See
if on of the display message
indicates any useful information.
If the display message reads
like the message above or the
one below, remove the cover
and reference the next page to
investigate further.