If the Oxygen Concentrator Pressure is
Low:
Remove the front cover to access the internal
adjustment knob of the regulator by extracting
the 4 phillips head screws.
Pull out the knob to engage the regulator &
adjust to 11.5psi on the right side external
gauge. The internal gauge above the knob will
fluctuate and should be ignored.
NOTE: if there
is a powdery
substance
all over the
internals the O2
Concentrator
has failed. Note
the internal
hour meter any
time the cover
is removed. Life
expectancy is
16,000 hours.
System Troubleshooting - Minor Fault
Oxygen Concentrator Fault
Verify the Flowmeter Setting & O2
Concentrator has no internal leaks.
Remove the line as shown in the picture on the
right. Adjust the black knob on the flowmeter
and set the ball to “6”.
Cover the output fitting with a finger and verify
the ball “crashes” to the floor of the flow meter.
If it does not - the concentrator has an internal
leak. If the flow meter ball (once the line is
reconnected) returns to 2-4 on the flowmeter,
all appears normal.
NOTE: if the ball in the flowmeter
does not drop
after reconnecting - this indicates that there is a
leak somewhere past the output and likely is in
the ozone generator itself.
Checking The Concentrator Adjustments
Pressure Switch Location
The device that generates the O2 Low error
is located in the ozone generator itself. It
can be found for further troubleshooting by
removing the cover of the ozone generator
and locating it behind the gauge panel.
Message on Ozone Generator Panel