MicroFID Software Manual
MicroFID. The filter will remove hydrocarbon contamination from
room air to provide zero grade air to MicroFID.
NOTE:
The charcoal filter does not filter methane or ethane. If these
compounds are present, you should use a gas bag with a supply of
commercial zero air.
To connect a charcoal filter:
1. Load the Teflon ferrules into the nut. The nut and
ferrules are supplied with the filter. See Figure 17.
2. Connect the nut to MicroFID’ s inlet. Do not tighten the
nut.
CAUTION
Overtightening the Teflon ferrules will result in damage to the
ferrules.
3. Remove the charcoal filter from its plastic bag and
insert it into the nut. Fingertighten the nut onto the
inlet. If the filter is not secure, ensure you have inserted
the tube stub far enough into nut. Do not overtighten
the fitting.
1/4" Nut
Connect
to MicroFID
Charcoal Filter
Front Ferrule
Back Ferrule
Figure 17
Connecting Charcoal Filter
96
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...