MicroFID Software Manual
4. Program and calibrate all the Cal Memories you will be
using. See Calibration on page 65. After calibration is
complete, sample the bag of calibration gas and the bag of
zero air to ensure MicroFID has been calibrated correctly.
5. Press the MAX key and clear the Max register if necessary.
6. If you are using an averaging interval, you may also want to
delete all events from the datalogger to avoid confusion
between different days’ data and to avoid running out of
space in the datalogger. See EVENT Key on page 50.
7. If you are performing Method 21 monitoring ensure you
have programmed and calibrated all the Cal Memories. You
must also program your monitoring schedule. See Method
21 Operation on page 72 for more information.