Appendices
Hydrogen cylinder
capacity:
9.2 liters at 1800 psig
Hydrogen cylinder
discharge time:
12 hours
Materials in
Stainless steel, Teflon®, Viton®,
sample stream:
polypropylene, PEEK, nitrile
chorobutadiene rubber
Inlet filter:
Replaceable 1 µm
Teflon/Polypropylene,
Inlet flow rate:
600 mL/min. +/ 10%
Operating
temperature range:
5° to 40°C (41° to 104°F )
Storage
temperature range:
5° to 40°C (41° to 104°F )
Operating
humidity range:
0 to 100% relative humidity (non
condensing)
Operating
0.5 to 2000 ppm methane equivalent
concentration range:
(Low Range), 10 to 50,000 ppm
methane equivalent (High Range)
Accuracy:
Methane: (after calibration with zero
air and 500 ppm methane span gas):
/0.5 ppm or +/10% of
actual methane concentration. (0.5
to 2000 ppm range)
Precision:
+/0.3 ppm or +/5% (0.5 to 2000
ppm methane range) and +/3 or +/
5% (10 to 50000 ppm methane
range)
Response time:
Less than 3 seconds (to 90% of full
scale response)
Detection limit:
0.5 ppm methane
145
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...