MicroFID Software Manual
Locking Ring
Tube Stub
Gas Bag Valve
Injection Port
Septum Retainer
Replacing the Septum
Remove retainer and
replace silicone septum.
Place septum with
Teflon side down.
Replace retainer
Do not overtighten.
Overtightening may result
in needle blocakge
Replace septum after
510 injections.To replace
septum, turn retainer counter
clockwise. For replacement
septa order Teflon bonded to
Silcone, diameter = .375" (9.5 mm)
Figure 11
Using the Gas Bag
8. Turn the regulator knob counter clockwise about half a
turn to start the flow of gas. Fill the gas bag about half
full and then close the regulator.
9. Disconnect the gas bag from the gas bag adapter and
empty the bag. Flush the bag a few times with the
calibration gas and then fill it.
10. Close the gas bag valve. Gently pull the valve tube
away from the bag, and then turn the knurled plastic
knob clockwise to tighten it against the valve tube.
11. Remove the knurled nut on the adapter tube from the
regulator.
12. Repeat this procedure, if necessary, to prepare a bag of
zero air.
NOTE:
Do not use the same gas bag or gas bag adapter for the bag of zero
air. You will contaminate the bag of zero air.
68
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...