MicroFID Software Manual
Baud
Rate
SW1
SW2
SW3
38400
On
On
On
19200
Off
On
On
9600
On
Off
On
4800
Off
Off
On
2400
On
On
Off
1200
Off
On
Off
600
On
Off
Off
300
Off
Off
Off
Table 5
DIP Switch Settings for Baud Rate
Parity
SW5
SW6
None
Off
Off
Odd
On
Off
Even
Off
On
Not Used Off
Off
Table 6
DIP Switch Settings for Parity
6. Switch 7 sets the Handshake mode. Off = Xon/Xoff. On
= DTR Mode. Set the Handshake mode to Xon/Xoff.
7. Use switch 8 to set the data flow direction to serial to
parallel. Off = Serial to Parallel. On = Parallel to Serial.
100
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...