![PerkinElmer Photovac MicroFID User Manual Download Page 50](http://html1.mh-extra.com/html/perkinelmer/photovac-microfid/photovac-microfid_user-manual_1530170050.webp)
MicroFID Software Manual
1. Press CAL and select the desired Cal Memory. Press ENTER
and then press EXIT. You do not want to calibrate MicroFID
at this point.
2. Press ALARM to display the current alarm level. If this value
is correct, press EXIT to return to the normal display.
3. If a new value is to be set, enter the new value, and press
ENTER. This alarm level is specific to the current Cal
Memory.
When an alarm condition is detected, the instrument status changes
to “A larm” and an audio signal is generated. The “Alarm” status
will not clear until the detected concentration falls below the set
alarm level.
PLAY Key
The PLAY key plays back previously recorded data.
To play back previously recorded data:
1. Press PLAY. Two options are available. Pressing ENTER
begins playback where it was last stopped. Press the SETUP
(*) key to set the playback options.
2. Select the start Event. MicroFID will display a list of available
events. Use the ARROW keys to scroll through the list. When
the start event you want is displayed, press ENTER.
NOTE:
An event name may not be available if it has been deleted.
3. If data has been averaged over an interval, select which value
is to be displayed, either the Minimum, the Average, or the
Maximum, and press ENTER. If you are playing back Method
21 data select Background, Sample or Difference to be
displayed.
4. The data can be played back in either numerical or bar graph
display. Press the DISPLAY key to select the desired option.
48
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...