MicroFID Software Manual
3. Remove the filter from its plastic bag and insert it into
the nut.
4. Fingertighten the nut onto the inlet. If the filter is not
secure, ensure you have inserted the tube stub far
enough into nut. Do not overtighten the fitting.
CAUTION
Overtightening the Teflon ferrules will result in damage to the
ferrules.
5. Calibrate all Cal Memories, that you are using, before
continuing operation.
The coalescing filter will remove particulates for up to 5 days of
operation. The exact duration will be determined by the operating
environment. MicroFID’ s inlet flow rate and sensitivity decrease
when the filer is full.
Serial to Parallel Converter
This device allows MicroFID, which is only capable of serial
communication, to communicate with a parallel device. This will be
most useful for printing, as most printers utilize parallel
communication. Your printer must have a print width of at least 65
characters and a proportionally spaced font.
DIP Switch Settings
The DIP switch settings are very important and must be set
correctly in order to correctly communicate with MicroFID.
MicroFID will always be the serial device so the data flow direction
will be serial to parallel.
98
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...