Accessories
Chart Recorder
MicroFID’s output can be displayed as a 0 to 1 volt analog voltage
on a chart recorder in real time.
NOTE:
MicroFID is not classified for use in hazardous locations with a
chart recorder.
To connect MircoFID to a chart recorder:
1. Set the chart recorder to 1 volt full scale.
2. Turn MicroFID off and connect the analog output cable
(Part No. MX396052) to the I/O port.
3. Connect the analog output cable to the chart recorder.
4. Turn MicroFID on.
The analog output has been fixed at 50000 ppm as full scale.
Gas Sampling Bag
A gas sampling bag is required for calibration of MicroFID. If you
are unsure of the quality of ambient air you should use an additional
gas bag filled with zero air. Connect the sample bag to the inlet
fitting with the gas bag adapter. See Calibration on page 65 for
details.
One sampling bag is included in the calibration kit. Additional gas
bags are also available (Part No. MX396017).
Charcoal Filters
A charcoal filter (Part No. MX396021 or MX396022) may be
connected to MicroFID during calibration to provide clean air to
95
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...