MicroFID Software Manual
Connecting the Refill Adapter to the
Hydrogen Tank
NOTE:
Read through this section before filling the internal cylinder.
Store the hydrogen tank in a well ventilated area, well away from
heat or possible ignition sources.
WARNING
Hydrogen gas is a fire and explosion hazard when exposed to heat or
flame. The lower explosive limit is 4%.
1. Connect the refill adapter to the tank of hydrogen. Hydrogen is
usually supplied with a CGA 350 cylinder valve outlet. The
refill adapter is supplied with the matching fitting for ease of
connection.
NOTE:
Do not force the connection.
Do not use Teflon tape with CGA fittings. In general, these fittings
are designed for metal to metal sealing.
Do not use adapters to connect one CGA fitting to another type of
CGA fitting. If the refill adapter does not match the outlet on your
hydrogen tank contact PerkinElmer.
2. Tighten the refill adapter onto the tank with a wrench. Do not
over tighten.
3. Hold the flexible hose and open the cylinder. Ensure there are
no leaks. Use soap solution to leak test the connection.
32
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...