Tutorial Session
Instrument Status
Detected Concentration
Time
Date
Event Name
Figure 4
Default Display
The default display provides the following information: instrument
status, current detected concentration, Event name (if the datalogger
is on), time, and date. If an event name is longer than three
characters the bottom line of the display will scroll through the
information.
The instrument status appears at the left of the upper line of the
display and on the Print and Graph outputs. Each status has a
priority assigned to it. If more than one status is in effect, then the
status with the highest priority is displayed until the condition is
corrected or until the option is turned off.
When the bar graph is selected, the instrument status is reduced to a
oneletter code.
Status
Code
Priority Description
NoFlm
N
1
Flame has not been ignited.
Check
K
2
One of 4 faults is occurring. Press the
TUTOR key for details.
Over
O
3
Detector electronics have been saturated.
Table 1
Instrument Status
41
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...