![PerkinElmer Photovac MicroFID User Manual Download Page 154](http://html1.mh-extra.com/html/perkinelmer/photovac-microfid/photovac-microfid_user-manual_1530170154.webp)
MicroFID Software Manual
Propylene Oxide
2.5
C
Styrene
1.2
L
1,1,2,2Tetrachloroethane
1.8
L
Tetrachloroethylene
(Perchloroethylene)
2.9
C
Toluene
0.9
C
1,1,1Trichloroethane
1.4
C
1,2,4 Trichlorobenzene
1.1
1,1,2Trichloroethane
1.7
L
Trichloroethylene (TCE)
2.8
C
Triethylamine
1.1
L
Vinyl Acetate
4.4
L
Vinyl Bromide
1.5
C
Vinyl Chloride (Chloroethylene)
2.1
C
Vinylidene Chloride (1,1DCE)
2.6
C
orthoXylene
1.1
L
metaXylene
1.2
L
paraXylene
1.2
L
Table 13
MicroFID Response Factors
Standards used for determination of these response factors were
derived from a variety of sources:
C
Certified gas cylinder, +/ 2% analytical accuracy
(Isobu/ 5% analytical accuracy)
G
Standard prepared by dilution of pure gas into zero air,
accuracy unknown
L
Standard prepared by addition of pure liquid to zero air,
accuracy unknown
152
Summary of Contents for Photovac MicroFID
Page 2: ......
Page 3: ...Photovac MicroFID Portable Flame Ionization Detector Part 396003 Rev G 11 ...
Page 5: ...Reference No 600757 Rev G 33 ...
Page 9: ...Notices and Warnings 77 ...
Page 10: ......
Page 17: ...Introduction 1 ...
Page 18: ......
Page 40: ...Tutorial Session 2 ...
Page 41: ......
Page 65: ...Detailed Operation 3 ...
Page 66: ......
Page 87: ...Accessories 4 ...
Page 88: ......
Page 108: ...MicroFID Software Manual 106 ...
Page 109: ...Routine Maintenance 5 ...
Page 110: ......
Page 115: ...Troubleshooting 6 ...
Page 116: ......
Page 136: ...Technical Description 7 ...
Page 137: ......
Page 143: ...Technical Description 141 ...
Page 144: ...Appendices 8 ...
Page 145: ......
Page 157: ...Appendices 155 ...