2
Chapter 2 Parameters and Features
2.1 Basic Performance Parameters
●
Horizontal
Linearity Error
≤
0.1%
●
Vertical Linearity
Error
≤
3%
●
Surplus Sensitivity
≥
62dB
●
Dynamic Range
≥
40dB
●
Electronic Noise
Level
≤
20%
●
Far-field
Resolution
≥
30dB
●
working Mode
Single-probe or Dual-probes
●
Transmit Pulse
400V negative spikes
●
Operating
Frequency Range
0.5MHz~20MHz, frequency subsection option, match
probes well
●
Detection Range
0.5mm~9999mm ( steel; longitudinal wave),
adjustable continuously, switchable steps
●
PRF
range
≤
125mm: 1KHz;
125mm<range
≤
1400mm: 250Hz;
1400mm<range
≤
9999mm: 40Hz
●
Sampling Rate
400MHz hardware real-time sampling
●
Velocity
1000m/s~15000m/s, adjustable continuously,
switchable steps
●
Display Delay
-10mm~1000mm, adjustable continuously, switchable
steps
●
Probe Delay
0
μ
s~199.9
μ
s
●
Damping
50
Ω
, 100
Ω
, 200
Ω
, 500
Ω
●
Gain Adjustment
0dB~110dB, step by 0.0/0.1/0.5/1.0/2.0/6.0dB
●
Rectification Mode Full wave / Positive half-wave / Negative half-wave /
RF
●
Display Screen
5.7" high brightness TFT color LCD screen, 640
×
480