16
5.1.2 PULS Menu
The main menu "PULS" contains one-page’s submenus. Refer to Table 5-2.
Table 5-2 PULS Menu
NO
Submenus
Details
1-1 DAMPING
50
Ω
/100
Ω
/200
Ω
/500
Ω
; the default is “50
Ω
”.
Change the amplitude/width/resolution of
echo by adjusting damping.
When damping is “high”, you will get higher
amplitude and wider echo width, but lower
resolution; and vice versa.
Press corresponding submenu selection key
or [+/-] key to switch damping value.
1-2 POWER
high/low; the default is “low”.
Adjust transmit power, chose "low" when
higher resolution is needed, while ''high" is
suitable for huge workpiece or through
transmission technique.
Press corresponding submenu selection key
or [+/
-
] key to switch between high and low.
1-3 DUAL
off/on; the default is “off”.
off: single element mode; on: double
elements mode.
Press corresponding submenu selection key
or [+/
-
] key to select working mode.
1-4 FREQ
BAND
0.5~22 /1~6 / 3~22MHz.
Chose appropriate frequency range of
amplifier according to probe frequency.
Press corresponding submenu selection key
or [+/
-
] key to select frequency band.