5.1.3 PROBE Menu
The main menu "PROBE” contains one-page’s submenus. Refer to Table 5-3.
Table 5-3 PROBE Menu
NO
Submenus
Details
1-1 TYPE
SPEC
Input specifications of probe: probe type+
fre crystal m element
size.
“Probe Type”: N (normal probe) / A (angle
probe)
Press corresponding submenu selection
key to change the background into yellow
among the submenus; press [+] key to
increase or [
-
] key to decrease the value.
1-2 P-DELAY
0.00~199.9
μ
s; the default is “0.00
μ
s”.
Set the zero point of probe.
Press corresponding submenu selection
key to switch between rough adjustment
and fine adjustment; press [+] key to
increase or [
-
] key to decrease the value.
1-3 ANGLE
Refraction angle: 0-85°.
Press corresponding submenu selection
key to switch between rough adjustment
and fine adjustment; press [+] key to
increase or [
-
] key to decrease the value.
Note: it is invalid when the normal
probe is selected.
1-4 X-VALUE
Distance between front edge of angle
probe and incidence point.
0.0~100.0mm; step by 0.1mm.
Press corresponding submenu selection
key to switch between rough adjustment
and fine adjustment; press [+] key to
increase or [
-
] key to decrease the value.
Note: it is invalid when the normal
probe is selected.
5.1.4 CAL Menu
The main menu "CAL” contains two-pages’ submenus. Refer to Table 5-4.