27
Table 5-13 MEAS MODE Menu
NO
Submenus
Details
1-1 TOF
Set measurement mode; peak/flank; the
default is "peak".
peak: select highest amplitude within the
gate as measurement point.
flank: select cross point between first echo
within the gate and gate itself as
measurement point.
Press corresponding submenu selection
key or [+/
-
] key to select measurement
mode.
1-2 RECTIFY
full wave /positive wave /negative wave /rf
Press corresponding submenu selection
key or [+/
-
] key to select rectify mode.
1-3 ZOOM
Used to magnify gate A; Gate A / off; the
default is "off".
Gate A: magnify gate A to entire echo
display area.
off: do not magnify Gate A.
Press corresponding submenu key or [+/
-
]
key to switch between "Gate A" and "off".
Note: this function can only be used
when Gate A logic is not "off".
1-4 ENVELOPE
on/off, the default is “off”.
on: enable envelope function.
off: disable envelope function.
Press corresponding submenu key or [+/
-
]
key to switch between "on" and "off".
5.1.14 MEAS DOT Menu
The main menu " MEAS DOT” contains one-page’s submenus. Refer to Table
5-14.
Table 5-14 MEAS DOT Menu
NO
Submenus
Details
1-1 MEAS-P1
Select measurement value to be displayed
at the lower side of the echo display area.
Options: Sa, Sb, Sb-a, Ha%, Hb%, Ha dB,
Hb dB, R-start, R-end, La, Lb, Da, Db, Pa,
Pb, Ra, Rb, Ha %crv, Hb %crv, DAC dB,
AVG dB, off.
Press corresponding submenu key to
select, then press [+/
-
] key to select
options.
1-2 MEAS-P2
Refer to “1-1” above.
1-3 MEAS-P3
Refer to “1-1” above.
1-4 MEAS-P4
Refer to “1-1” above.