26
Table 5-12 b GAT Menu
NO
Submenus
Details
1-1 b
LOGIC
Used to select Gate B working mode and
alarm logic; off/positive/negative; the
default is “off”.
off: Gate B un-visible, alarming and
measurement function can not be used.
positive: if echo amplitude within Gate B is
higher than the gate threshold, alarm light
indicator will be on and there will be sound
alarm when "Horn" is set to "on".
negative: if echo amplitude within Gate B is
lower than the gate threshold, alarm light
indicator will be on and there will be sound
alarm when "Horn" is set to "on".
Press corresponding submenu selection
key or [+/
-
] key to select working mode of
Gate B.
1-2 b
START
0.00~9999mm; the default is “14.00mm”
when “positive” or “negative” is selected.
Press corresponding submenu selection
key to switch between rough adjustment
and fine adjustment; press [+] key to
increase or [
-
] key to decrease the value.
1-3 b
WIDTH
0.020~9999mm; the default is “10.00mm”
when “positive” or “negative” is selected.
Press corresponding submenu selection
key to switch between rough adjustment
and fine adjustment; press [+] key to
increase or [
-
] key to decrease the value.
1-4 b
THRSH
5%~90%; step by 1%; the default is “50%”
when “positive” or “negative” is selected.
Press corresponding submenu key to
select; press [+] key to increase or [
-
] key
to decrease the value.
5.1.13 MEAS MODE Menu
The main menu " MEAS MODE” contains one-page’s submenus. Refer to Table
5-13.