4.4 Probe Connecting
Connectors of Instrument are standard Lemo-01. Operating in single element
method, both sockets of connector are applicable (parallel connection inside).
When connecting the double elements probes or two probes for receive-transmit
work, please make sure to connect the transmitter probe to the right socket (“ ”
mark on the instrument crust), and the receiver probe to the left one (“ ” mark
on the instrument crust). Wrong connection may cause unsatisfied test effects.
4.5 Power On/Off
The basic operation for powering on/off is as follows.
(1)
Powering on:
long press [Power] key to start up the instrument.
(2)
Self-checking after Start-up:
the instrument will complete the Self-checking
automatically after start-up, then it will enter the previous state with the same
parameters keeping when normally powered off.
(3)
Checking battery voltage:
please check the power monitor icon in the lower
right corner of the screen at times. If the power is insufficient, the instrument
will power off automatically after alarming for one minute.
(4)
Working:
set the instrument parameters or do the testing.
(5)
Powering off:
press [Power] key for some time to shutdown the instrument
normally, the current settings are saved.
Note: it is in the abnormal shutdown state when the power is cut off
suddenly while testing, the current settings may be not saved.
4.6 Software Update and Return to Factory Default Settings
The instrument has many updating methods as follows, which are convenient for
on site detection.
(1)
Software update
: in the condition of start-up, long time press [PgUp] and
[PgDn] keys at the same time and keep still, according to kind reminders on
screen, inset USB disk with new edition process, press [
] key which is
corresponded to the red words "Ver 1.10", then software update prompt will
display on screen , after finishing, press [PgUp] and [PgDn] keys at the same
time once, instrument returns to the normal detection condition (but it will not
change the stored data which has been saved in the STORE menu).
(2)
Parameters
being
returned to factory settings
: press [Power] key to start