15
Chapter 5 Menu Introduction
5.1 Normal Menus
PCE-USC 20
contains 17 main menus (in 4 pages), each main menu contains
several submenus, and the specific structure, parameter setting and basic
operations are as follows.
5.1.1 BASE Menu
The Main Menu "BASE" contains one-page’s submenus. Refer to Table 5-1.
Table 5-1 BASE Menu
NO
Submenus
Details
1-1 RANGE
“Detection Range”: 0.500~9999mm (steel
longitudinal wave); the default is “100mm”.
Press corresponding submenu selection
key to switch between rough adjustment
and fine adjustment; press [+] key to
increase or [
-
] key to decrease the value.
1-2 MTLVEL
“Material Velocity”: 1000~15000m/s; the
default is 5920m/s.
Press corresponding submenu selection
key to switch between rough adjustment
and fine adjustment; press [+] key to
increase or [
-
] key to decrease the value.
1-3 D-DELAY
“Display Delay”: -10~2000mm; the default
is 0.00mm.
Press corresponding submenu selection
key to switch between rough adjustment
and fine adjustment; press [+] key to
increase or [
-
] key to decrease the value.
1-4 THICKNESS
“Thickness Range”: 1~9999mm.
It is invalid if the normal probe N is
selected.
It is used to set the workpiece thickness of
the detection surface vertically if the angle
probe A is selected, and the default is
100mm.
Press corresponding submenu selection
key to switch between rough adjustment
and fine adjustment; press [+] key to
increase or [
-
] key to decrease the value.