22
5.1.8 AUTO SETUP Menu
The main menu "AUTO SETUP” contains one-page’s submenus. Refer to Table
5-8.
Table 5-8 AUTO SETUP Menu
NO
Submenus
Details
1-1 AUTO
GAIN
Set amplitude of echo within gate A to
desired height (can be adjusted).
Press corresponding submenu key or [+]
key to adjust the value.
1-2 AGC
20%~80%; step 1%; the default is “50%”.
Set amplitude of "auto gain".
Press corresponding submenu key to
select; press [+] key to increase or [
-
] key
to decrease the amplitude value.
1-3 SURF
AMEND
¢10~¢2000mm or flat; the default is “flat”.
Press corresponding submenu key to
select or switch step; press [+] key to
increase or [
-
] key to decrease the
amplitude value.
It is invalid when the normal probe is
selected.
1-4 PEAK
HOLDING
off/on; the default is “off”.
on: A purple "×" will be displayed on top of
the maximum echo inside the gate, if a
new maximum echo appears in the gate,
then the "×" will move to indicate the
position of the new one, otherwise the "×"
remains.
off: this function can not be used.
Press corresponding submenu key or [+/
-
]
key to switch between "on" and "off".