45
parameters of current channel is recalled.
(5) Deletion: Select DELETE submenu, press [+] key, then stored detect
parameters of current channel is deleted.
Note: as to "SAVE/DELETE" submenu, the default is "DELETE" when there
be stored parameters in DATA ID; otherwise the default is "SAVE".
6.12 Data Store and Replay
6.12.1 Summarize
Lots of echo data need to be recorded in flaw detection process, users can keep
required echo and data in the instrument as important content of flaw detection
report. The data may be directly replayed in the instrument for viewing, also can
be kept in computer for unified store through communication interface.
6.12.2 Operation Steps
(1) Enter STORE main menu, select the top submenu DATA ID, menu interface as
follows. Refer to Figure 6-17.
(2) Press [+/
-
] key to select DATA ID.
Figure 6-17 Data Store and Replay
(3) Store: Select SAVE sub-menu, press [+] key to confirm data storage. In front of
the channel with storage data has “*” mark. On the contrary, it does not have
“*” mark for blank channel.
(4) Recall: Select DATA RECALL sub-menu, press [+] key, then stored waveform