Chapter 4 Operating Instruction
4.1 Power Supply
Digital ultrasonic flaw detector
3&(86&
can be operated by use plug-in unit
(AC adapter) or batteries. Put battery into instrument, connect the instrument to
adapter for power supply also charge battery.
4.2 Operating Instruction of the AC Power Supply
The instrument can use the AC power to supply power by being connected to the
special AC adapter.
Note: in the condition of the instrument using the AC adapter for power
supply without loading the battery, if shut it off forcibly, the
instrument is under abnormal shutdown state and the parameter
previous settings may not be saved. The correct way to shutdown is
to long press the power key for about 2s.
4.3 Battery Power Supply
Using battery for power supply, please use batteries manufactured for matching
instrument from factory.
(1)
Insert battery:
battery compartment is in the back side of instrument, use
screwdriver to loosen the cover of battery compartment, put battery into
compartment, close the battery compartment then screw down bolts.
(2)
Battery power indicator:
there are five grids at lower right corner of display
screen to show power supply condition, if these grids are less then two, stop
flaw detection immediately, save test parameters and data, replace battery or
charge battery promptly.
(3)
Charge battery:
please use power adapter manufactured for matching
instrument from factory. Before use adapter, please read instruction carefully. 5
hours required for battery continual charging.
Note: if long-time detecting is needed on site, please bring the reserve
battery with you. You should pay more attention to it especially in
winter when the temperature is low outside and the efficiency of
battery discharging decreases.