23
Table 5-9 WELD GRAPH Menu
NO
Submenus
Details
1-1 WELD
TYPE
Select groove type: V/X/off.
Press corresponding submenu key or [+/
-
]
key to select groove type.
It is invalid when the normal probe is
selected.
1-2 P-PLACE
1~500mm; step by 1/10/100mm; the
default is “100mm”.
Input the distance between probe front
edge and weld center line.
Press corresponding submenu selection
key to switch step; press [+] key to
increase or [
-
] key to decrease the value.
1-3 WELD
WIDTH
1~200mm, step by 1/10mm; the default is
“100mm”.
Input the width of groove.
Press corresponding submenu selection
key to switch step; press [+] key to
increase or [
-
] key to decrease the value.
1-4
REFRESH
√
Press to generate the flaw position
indication diagram.
Press corresponding submenu key to
select; press [+] key to confirm refresh.
5.1.10 STORE Menu
The main menu "STORE” contains two-pages’ submenus. Refer to Table 5-10.
Table 5-10 STORE Menu
NO
Submenus
Details
1-1
CHANNEL ID
(See the next page)
The channel ID contains 1-30 ID
,
test
parameters will be stored in the selected
channel.
Press corresponding submenu selection
key to switch between “CHANNEL ID" and
"DATA ID"; press [+] key to increase or [
-
]
key to decrease the number.