5.1.16 CONFIG 1 Menu
The main menu " CONFIG 1” contains one-page’s submenus. Refer to Table 5-16.
Table 5-16 CONFIG 1 Menu
NO
Submenus
Details
1-1 LANGUAGE
Set language of the instrument; Chinese /
English; the default is “Chinese”.
Press corresponding submenu key or [+/
-
]
key to switch between languages.
1-2 UNIT
Set measuring unit: mm/inch; the default is
"mm".
Press corresponding submenu key or [+/
-
]
key to switch between units.
1-3 HORN
Set alarm mode: alarm/off; the default is
"off".
alarm: sound alarm will act if it satisfies the
alarm condition.
Press corresponding submenu key or [+/
-
]
key to switch between "alarm" or "off".
1-4 RESTRAINT
Range: 0%~50%; the default is “0%”.
To eliminate unwanted waves, such as
structural noise from detected material or
other clutter waves. Any wave which is
lower than restraint value will be
eliminated, else remains.
Press corresponding submenu key to
select; press [+] key to increase or [
-
] key
to decrease the value.
Note: be cautious to use this function,
for it may eliminate echoes from
flaws. Thus it is not permitted to
use in many detection standards.