Headphones
The headphones are in constant contact with the patient, and should therefore be kept clean.
Clean the headphones between patients, e.g. with a non-alcohol based antibacterial wipe, such as Audiowipes.
Eartips for Insert Earphones
Warning
•
To prevent cross-infection, use new eartips when you test the next client.
The eartips are disposable and therefore should not be cleaned or re-used. There are no special requirements for the dis-
posal of the eartips.
Bone oscillator
Clean the bone oscillator between patients, e.g. with a non-alcohol based antibacterial wipe, such as Audiowipes.
10.3
Calibration
You can download new software to MADSEN Xeta, and if this involves no change to the transducers, MADSEN Xeta does
not need recalibration.
Annual calibration
The audiometer, headphones, bone oscillators, and sound field speakers must be calibrated once a year by your authorized
service department.
The audiometer is dispatched from the factory together with a Test Report (Calibration Certificate). The Test Report spe-
cifies the transducers that have been calibrated (i.e., those which have been supplied together with the instrument),
according to which standards, and the equipment used for calibration. Results are listed for each transducer at all standard
frequencies.
In general, the instrument is calibrated in dB HL and dB masking level using the stated reference equivalent thresholds; dB
HL is related to sound pressure levels, dB SPL = dB re 20
µ
Pa, and force levels (dB re 1
µ
N).
Caution
•
Note that calibration has been performed only on the transducers supplied! If you wish to use any other
transducer for testing with the device, please contact your local distributor first.
82
Otometrics - MADSEN Xeta
10 Maintenance and calibration
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...