Auto Threshold
Selects the Automatic Threshold test. For a description, see
Automatic Air Conduction testing
When activated, the button flashes to indicate that you have selected the test, or that the test has been
paused by the user, by faulty responses or by talk-over.
Auto Screening
Selects the Automatic Screening test. For a description, see
.
When activated, the button flashes to indicate that you have selected the test, or that the test has been
paused by the user, by faulty responses or by talk-over.
More (tests)
Press to see the optional test types. Press the desired function key to select.
For a description of the additional tests, see
•
SISI (Short Increment Sensitivity Index)
•
•
ABLB (Alternate Binaural Loudness Balance)
These tests are optional. To add any of these tests to your configuration, contact your local supplier.
3.2.5
Signal indicators
The indicators are located on either side of the display. They light up when the selected test signal is presented
in the corresponding ear.
•
Discreet mode
In the Auto Threshold and Auto Screening tests you can change the setting of the signal indicators to
Dis-
creet mode
, which means that they do not light up when the test signal is presented. If MADSEN Xeta is in
view of the patient, this will prevent the patient from knowing when a signal is presented.
This mode also prevents the patient from simulating hearing, because the stimulus level is hidden during
the automatic test.
3.2.6
Extended Range
Enables presentation of stimuli above the safety limit (approx. 100 dB). The safety limit is set to 20 dB
below the maximum dB HL for a given transducer and frequency.
Note
•
Automatic testing is always limited to 100 dB HL.
Extended Range
has 2 modes (to select, see
):
•
Auto
All levels at all frequencies are always accessible, but the Extended Range button will start flashing when a level above
100 dB HL is reached.
•
Manual
The first time extended range above 100 dB HL is reached after patient change or ear change, press the
Extended
Range
button
once
to allow for access to any level at any frequency with any selected manual test and signal type
Otometrics - MADSEN Xeta
19
3 Overview of MADSEN Xeta
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...