Bone
Selects bone conduction (via bone conductor on mastoid).
Reverse
Reverses the function of the
Present stimulus
button:
•
Active
(indicator lit)
The test signal is presented continuously and the stimulus is interrupted when you press the
Present stim-
ulus
button.
•
Inactive
(indicator not lit)
The test signal is presented when you press the
Present stimulus
button.
Note
•
The presentation only applies to the stimulus signal. The masking signal is normally always ON. To change the
masking interrupter's function see
.
Warble
Frequency modulates the stimulus signal ±5% at a rate of 5 Hz around the selected frequency.
Pulsed
The stimulus signal is presented with a Pulse/Pause ratio of 50%.
Timed
The stimulus signal is presented with a preset duration. Does not apply in
Reverse
mode.
3.2.3
MASKING
•
By default, you always adjust the stimulus signal level on the
Left Level
knob, regardless of whether you are testing
the right or left ear.
•
To adjust the masking intensity use the
Right Level
knob.
To change the default setting, see
Contra Air
Contralateral masking via stereo Air Conductor (headset or insert phones).
Mono Insert
Contralateral masking via mono insert phone (for bone conduction test only).
Level Lock
Makes the masking level follow the stimulus signal level. You can adjust the difference between the stimulus
signal and the masking signal by using the
Right level
knob.
3.2.4
TEST
Manual Tone testing is active by default.
18
Otometrics - MADSEN Xeta
3 Overview of MADSEN Xeta
Summary of Contents for Madsen Xeta
Page 1: ...MADSEN Xeta Reference Manual Doc No 7 50 0870 EN 17 Part No 7 50 08700 EN ...
Page 6: ...Index 103 6 Otometrics MADSEN Xeta ...
Page 26: ...3 Overview of MADSEN Xeta 26 Otometrics MADSEN Xeta ...
Page 44: ...4 Navigating in the OTOsuite Audiometry Module 44 Otometrics MADSEN Xeta ...
Page 54: ...5 Preparing for testing 54 Otometrics MADSEN Xeta ...
Page 66: ...6 MADSEN Xeta Test Settings 66 Otometrics MADSEN Xeta ...
Page 86: ...11 Unpacking and installing 86 Otometrics MADSEN Xeta ...
Page 88: ...12 Maximum non destructive voltage 88 Otometrics MADSEN Xeta ...
Page 94: ...14 Standards and safety 94 Otometrics MADSEN Xeta ...
Page 102: ...15 Technical specifications 102 Otometrics MADSEN Xeta ...